Model Number 16-02-85
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 01/27/2016
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. This event was initially considered to be non-reportable. However, after additional evaluation, livanova (b)(4) has decided to reclassify this event as reportable. This report is being filed as part of a retrospective review conducted in response to this new decision. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and traced it back to three defective heating elements. The unit was recommended to be returned to livanova (b)(4) for further investigation and repair. During investigation, one of the heating elements was determined having an insulation fault due to corrosion. To resolve the reported issue, all three heating elements were replaced. A functional control and a test run were performed and passed with no further issue shown. The device was returned to service. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event Description
Livanova (b)(4) received a report that during cleaning, the heater cooler system 3t kept tripping the fuses on the wall plug. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7106308
MDR Text Key94821036
Report Number9611109-2017-00996
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052601
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Other
Type of Report Initial
Report Date 12/11/2016
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number16-02-85
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2016
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/27/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-