• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT G3+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT G3+ CARTRIDGE Back to Search Results
Catalog Number 03P78-50
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Atherosclerosis (1728)
Event Date 11/24/2017
Event Type  malfunction  
Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2017, abbott point of care was contacted by a customer regarding i-stat g3+ cartridges that yielded a suspected discrepant result on a (b)(6) year old male patient with atherosclerosis.There was no additional patient information at the time of this report.Return product is not available for investigation.(b)(6).There are no injuries associated with this event.The investigation is underway.
 
Manufacturer Narrative
(b)(4).The investigation was completed on 01/26/2018.Retain product was tested and functioning according to specification.Return product was not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT G3+ CARTRIDGE
Type of Device
G3+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE CANADA LTD.
185 corkstown road
ottawa, ontario K2H 8 V4
CA   K2H 8V4
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key7106857
MDR Text Key94831510
Report Number2245578-2017-00632
Device Sequence Number1
Product Code CHL
UDI-Device Identifier10054749002280
UDI-Public10054749002280
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K940918
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2018
Device Catalogue Number03P78-50
Device Lot NumberD17224
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-