The fse replaced the cam follower on site and will not be returned to isi for evaluation as it was scrapped in the field.A follow-up mdr will be submitted if additional information is received.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if the reported malfunction were to recur if could cause or contribute to an adverse event.This report does not admit that the report or information submitted under this report constitutes an admission that the device, intuitive surgical or intuitive surgical employees, caused or contributed to the reportable event.
|
It was reported that during a da vinci-assisted myomectomy procedure, the surgeon was unable to turn the patient side cart (psc) left or right to dock to the patient.The intuitive surgical, inc.(isi) technical support engineer (tse) instructed the customer to put the psc drive in neutral position, however, the issue persisted.At that time, the surgeon made the decision to abort the case post anesthesia and port placement.There was no report of patient harm, adverse outcome or injury.An isi field service engineer (fse) was dispatched to the facility and was able to reproduce the reported failure.The fse replaced the cam follower to resolve the issue.The cam follower steering linkage connects the left/right steering linkages to the tiller on the psc.
|