Model Number IS3000 A60P9B
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2017
Event Type  malfunction  
Manufacturer Narrative
The fse replaced the cam follower on site and will not be returned to isi for evaluation as it was scrapped in the field.A follow-up mdr will be submitted if additional information is received.This complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if the reported malfunction were to recur if could cause or contribute to an adverse event.This report does not admit that the report or information submitted under this report constitutes an admission that the device, intuitive surgical or intuitive surgical employees, caused or contributed to the reportable event.
 
Event Description
It was reported that during a da vinci-assisted myomectomy procedure, the surgeon was unable to turn the patient side cart (psc) left or right to dock to the patient.The intuitive surgical, inc.(isi) technical support engineer (tse) instructed the customer to put the psc drive in neutral position, however, the issue persisted.At that time, the surgeon made the decision to abort the case post anesthesia and port placement.There was no report of patient harm, adverse outcome or injury.An isi field service engineer (fse) was dispatched to the facility and was able to reproduce the reported failure.The fse replaced the cam follower to resolve the issue.The cam follower steering linkage connects the left/right steering linkages to the tiller on the psc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DA VINCI SI SURGICAL SYSTEM
Type of Device
ENDOSCOPIC INSTRUMENT CONTROL SYSTEM
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 
4085232420
MDR Report Key7108974
MDR Text Key95773509
Report Number2955842-2017-00834
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081137
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIS3000 A60P9B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-