Model Number S7 |
Device Problems
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient age not available from the site.
Patient weight not available from the site.
A medtronic representative went to the site to test the equipment.
The representative reported that the polaris spectra system control unit (scu) to positioning sensor unit (psu) cable was replaced.
The hardware, software, and instruments passed the system checkout.
The system was found to be fully functional.
The scu to psu cable was returned to the manufacturer for analysis.
The device was found to be fully functional with no problem found.
The reported event could not be duplicated by medtronic personnel.
|
|
Event Description
|
A medtronic representative reported that, while in a spinal fusion, the camera of the navigation system lost functionality.
After manual manipulation of the camera cable, functionality was restored.
There was a reported delay to the procedure of less than 1 hour due to this issue.
There was no impact on patient outcome.
No additional information was provided.
|
|
Manufacturer Narrative
|
Additional information: patient age and weight provided.
|
|
Manufacturer Narrative
|
Patient identifier field not sufficient to hold all digits, should read: (b)(6).
|
|