Model Number FUSION
Device Problems Computer Software Problem (1112); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/14/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. It was reported that the computer for the navigation system was replaced. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional. The suspect computer for the navigation system has been received by the manufacturer for evaluation. However, results are not available at this time.
 
Manufacturer Narrative
The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided. The analysis of the computer for the navigation system was completed by medtronic personnel. The device was found to be fully functional with no problem found. The reported event could not be duplicated by medtronic personnel.
 
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. No parts have been received by the manufacturer for evaluation. No parts have returned for analysis.
 
Event Description
A medtronic representative reported that outside of a procedure, upon boot-up, the navigation system never reached the application launch screen; all that was shown was linux lines. Restarting the system did resolve the issue, and the system started working again. There was no patient present when this issue was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7110226
MDR Text Key103037045
Report Number1723170-2017-05083
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/12/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-