Model Number FUSION
Device Problems Computer Software Problem (1112); Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/14/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. A medtronic representative went to the site to test the equipment. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional. Software analysis was completed and the issue was suspected to be caused by faulty media as the representative confirmed the site was able to replicate the issue with the disc. It was possible to load other discs without issue. No parts have been returned to the manufacturer for evaluation.
 
Event Description
A site representative reported that outside of a procedure, the navigation system was exiting. The site had loaded a disc and after sitting on the select patient screen for approximately 5 minutes, the system exited to the blue screen. The site rebooted the system and were able to get to the same point again, however the issue repeated twice more. It was reported that it was possible to load other discs without issue. The issue was replicated with the original disc. There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stephanie riley
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7110255
MDR Text Key94965308
Report Number1723170-2017-05084
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 12/12/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/15/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-