• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.240
Device Problem Component Missing (2306)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2017
Event Type  malfunction  
Manufacturer Narrative
Reporter's number (b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the recon sagittal saw device was missing a screw.It was unknown if there were any delays to the surgical procedure or if a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device coupling tool side did not function.The device also failed pretest for check falling out protection (steal ring), check the saw head¿s locking mechanism and check the saw blade coupling.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper device maintenance.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECON SAGITTAL SAW WITH KEY FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 
6107195000
MDR Report Key7114381
MDR Text Key95058415
Report Number8030965-2017-50614
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819462588
UDI-Public(01)7611819462588(11)130507
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-