Catalog Number UNKNOWN |
Device Problems
Difficult to Remove (1528); Unintended Movement (3026); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Vessel Or Plaque, Device Embedded In (1204); Pulmonary Embolism (1498); Myocardial Infarction (1969); Internal Organ Perforation (1987); Perforation of Vessels (2135); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Catalog# is unknown but referred to as cook celect filter.Name and address for importer site: (b)(4).Since catalog# is unknown 510(k) could be either k061815, k073374, k090140, k112119, k121057 or k121629.(b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2014".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
Exemption number e2016032e lot.(b)(4).Additional information: investigation : it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'celect, pe, tilt, unable to retrieve (emb'd), v.C.Organ perf, heart attack, kidney injury, anxiety, stress '.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.A filter that is embedded in the wall of the ivc may be difficult to retrieve.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Pe is a known risk in relation to filter implant reported in the published scientific literature.Also, it is reported that the pulmonary embolism in some cases may originate from upper extremities instead of lower extremity veins.With all filters, there is some risk of further pulmonary embolism.Unknown if the reported heart attack, kidney injury, anxiety, stress is directly related to the filter and unable to identify corresponding failure mode(s) at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Event Description
|
"ithis" additional information received on 13feb2018 as follows: ¿[pt] allegedly received an implant on (b)(6) 2014 via the right common femoral vein due to deep vein thrombosis with gastrointestinal bleeding and hypotension.[pt] is alleging tilt, device unable to be retrieved, organ perforation-mesenteric, embedment.[pt] further alleges anxiety/stress, post-implant pulmonary embolism.Filter retrievals were attempted on (b)(6) 2015 and (b)(6) 2015 and were unsuccessful.During the first attempt on (b)(6) 2015 pt developed atrial fibrillation approximately midway through the procedure with mild decrease in systolic blood pressure.
|
|
Manufacturer Narrative
|
Exemption number e2016032e lot.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
|
Manufacturer Narrative
|
Exemption number e2016032e lot.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(6).Additional information: investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'celect, pe, tilt, unable to retrieve (hook emb'd), organ perf, anxiety, stress'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.A filter that is embedded in the wall of the ivc may be difficult to retrieve.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Pe is a known risk in relation to filter implant reported in the published scientific literature.Also, it is reported that the pulmonary embolism in some cases may originate from upper extremities instead of lower extremity veins.With all filters, there is some risk of further pulmonary embolism.Unknown if the reported anxiety, stress is directly related to the filter and unable to identify corresponding failure mode(s) at this time.Rpn and lot# are unknown, but the celect filter is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Patient code: myocardial infarction ((b)(4)) not listed in the ifu,vessels, perforation of ((b)(4))-listed in ifu.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Patient code: myocardial infarction (1969) not listed in the ifu,vessels, perforation of ((b)(4))-listed in ifu.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
This additional information received on (b)(6) 2018 as follows: [pt] additionally alleges vena cava perforation, non st-segment elevation myocardial infarction w/demand ischemia & acute kidney injury 2/2 pe.
|
|