MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.201
Device Problem Device Operational Issue (2914)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Reporter's phone number: (b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event Description
It was reported from (b)(6) that the battery handpiece device was not working in reverse and making a noise.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device failed pretest for check status of development, check proper function of the triggers, and check of free moving.It was noted that the bearings were not function and were defective on the device and the reverse trigger was faulty.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg
SZ  
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 
6107195000
MDR Report Key7120081
MDR Text Key95196504
Report Number8030965-2017-50712
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819977815
UDI-Public(01)7611819977815(11)131025
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
EXMEPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-