Catalog Number UNKNOWN |
Device Problems
Unintended Movement (3026); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Internal Organ Perforation (1987); Perforation of Vessels (2135); Anxiety (2328); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Blank fields on this form indicate the information is unknown or unavailable.Catalog#: unknown but referred to as a cook günther tulip filter.(b)(4).Since catalog# is unknown the 510(k) could be either k090140, k112119 or k121057.(b)(4).It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating, ¿gunther tulip filter implanted".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "patient received a cook gunther tulip filter on (b)(6) 2012".It is alleged that patient was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).Additional information: investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tilt, vc and organ perforation, anxiety'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Unknown if the reported anxiety is directly related to the filter and unable to identify corresponding failure mode(s) at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).Corrected data: adverse event to adverse event and product problem.(b)(4).The event is currently under investigation.A supplemental report will be provided upon conclusion.
|
|
Event Description
|
This additional information receive d on 31jan2018 as follows; pt allegedly received an implant on (b)(6) 2012 via the right internal jugular vein due to deep vein thrombosis.Pt is alleging tilt, vena cava perforation, organ perforation (lumbar vein).Pt further alleges anxiety.
|
|
Event Description
|
No additional information provided at this time.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer) (b)(6).Registration no.: (b)(4).Additional information: investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tulip, vc + organ perforation, tilt, anxiety, mental anguish, stress'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Unknown if the reported anxiety, mental anguish, stress is directly related to the filter and unable to identify corresponding failure mode(s) at this time.Catalog and lot# are unknown, but the filter tulip is manufactured and inspected according to manufacturing instructions.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
This additional information received on 24may2018 as follows: pt alleges mental anguish, stress.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of (b)(4) (importer).G1) name and address for importer site: (b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|