MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM; EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION COMPACT
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.No parts have been received by the manufacturer for evaluation.Part not returned for analysis.
 
Event Description
A medtronic representative reported that, while in a functional endoscopic sinus surgery (fess), the navigation system displayed a "localizer was disconnected" error message.Restarting the navigation system did not restore functionality.Once the system was powered down, the power/communication cable for the axiem was disconnected, the navigation system restarted and the cable re-attached.The navigation system functioned as designed.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no reported impact on patient outcome.No additional information was provided.
 
Manufacturer Narrative
Correction: unique device identification (udi) and device manufacture date updated to proper value.
 
Manufacturer Narrative
Additional information: a medtronic representative reported that there were no procedures performed on the event date utilizing the medtronic navigation system.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7120960
MDR Text Key94999793
Report Number1723170-2017-05119
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 01/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Is the Reporter a Health Professional? No
Date Manufacturer Received01/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-