Patient information not provided due to (b)(6) patient privacy regulations.No procode, common device name, unique device identification(udi) and/or 510k provided as this device is not released for distribution in the united states.A medtronic representative went to site to test the equipment.Representative reported that when the bone model was used to verify accuracy, an inaccuracy was observed.Re-calibrating the system resolved the issue.No parts were replaced.No parts have been received by the manufacturer for evaluation.
|
A site representative reported that during spinal fusion procedure, during an accuracy check after 3d arcadis link was performed using a spine reference frame light, an imprecision of approximately 2 millimeters was observed on the anterior posterior (ap), si and lr directions.It was reported that when flouronav reference frame long cable was used to perform 3d arcadis link, less error than the initial arcadis was observed and the accuracy improved from the initial.The procedure was completed with the use of navigation.There was no delay to procedure.No impact on patient outcome.
|