• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Stops Intermittently (1599)
Patient Problem No Patient Involvement (2645)
Event Date 11/22/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that a computer replaced to resolve the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect computer has not been received by the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that while outside of procedure, the screen of the navigation system would intermittently freeze.There was no patient present at the time of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7123842
MDR Text Key95079373
Report Number1723170-2017-05183
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560
Is the Reporter a Health Professional? No
Date Manufacturer Received11/22/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-