Model Number 72200616
Device Problems Component Falling (1105); Leak/Splash (1354); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation. There was no relationship found between the returned device and the reported incident. A visual inspection was performed on the product and no issue was observed. No rusty water or loose debris was observed during visual inspection. Spigot and fork for blade attachment were free of contamination of any kind. A functional evaluation revealed a motor stall error and overheating. The complaint was not confirmed. A root cause for the seized motor has been associated with a corroded motor/gearbox. Factors that could have contributed to the event include cleaning and sterilization methods and the chemicals involved over a period of time.
 
Event Description
It was reported that during surgery with autolog hamstrings graft, debris and rust colored water was leaking from the shaver blade after approx. 4 min use. Shaver blade was removed from handpiece and a ¿donut¿ like substance with debris fell out from the handpiece. A rinse of the knee was done and as a safety precaution the patient was treated with antibiotics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7124544
MDR Text Key141497300
Report Number1643264-2017-01829
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/01/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-