Catalog Number UNKNOWN |
Device Problems
Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Internal Organ Perforation (1987); Perforation of Vessels (2135); Anxiety (2328); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'gunther tulip - vc & organ perf, anxiety - updated sfc'.
Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.
F.
R.
803.
56.
Vena cava wall perforation is a known potential complication of vena cava filters.
Both symptomatic and asymptomatic events have been reported.
Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.
G.
, a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.
There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.
E.
G.
Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.
In contrast, perforation of adjacent organs is reported with clinical sequelae.
Unknown if the reported anxiety is directly related to the filter and unable to identify corresponding failure mode(s) at this time.
Rpn and lot# are unknown, but the tulip filter is manufactured and inspected according to specifications.
No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.
G.
Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
Investigation - it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿gunther tulip filter implanted".
Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.
F.
R.
803.
56.
No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.
G.
Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
(b)(4).
This report includes information known at this time.
A follow up report will be submitted should additional relevant information become available.
|
|
Event Description
|
This additional information received on 13feb2018 as follows: pt allegedly received an implant on (b)(6) 2007 via the [right internal jugular vein for prevention of pulmonary embolism.
Pt is alleging vena cava perforation, organ perforation, perforation to the mesentery.
Pt.
Further alleges anxiety.
There were no reported device retrieval attempts.
|
|
Manufacturer Narrative
|
The event is currently under investigation.
A supplemental report will be submitted upon completion.
|
|
Event Description
|
It is alleged that the plaintiff received a gunther filter on (b)(6) 2007.
It is alleged that the plaintiff was injured without further explanation.
Hospital and medical records have been requested but not yet provided.
|
|
Search Alerts/Recalls
|