COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Catalog Number NGE-022115
Device Problems Break (1069); Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2016
Event Type  malfunction  
Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event Description
A left ureterorenoscopy was performed.A laser was used on the renal calculus and the basket to extract the stones.The complainant reported that the device would not work as the basket kept getting caught in cladding on the basket tip.It was further noted that the bronze lining on basket loosened and the basket appeared broken; the basket did not open and close as expected.The sheaths on the basket wires had been scraped.There were no reported adverse effects to the patient or additional procedures required as a result of this product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key7125120
MDR Text Key95828756
Report Number1820334-2017-04113
Device Sequence Number1
Product Code FFL
UDI-Device Identifier00827002482975
UDI-Public(01)00827002482975(17)181118(10)6390905
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNGE-022115
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
Patient Weight84
-
-