• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-FEM
Device Problems Fracture (1260); Difficult to Remove (1528); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Vessel Or Plaque, Device Embedded In (1204); Dyspnea (1816); Internal Organ Perforation (1987); Thrombosis (2100); Perforation of Vessels (2135); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
The event is currently under investigation.A supplemental report will be submitted upon completion.
 
Event Description
It is alleged that the plaintiff received a gunther filter on (b)(6) 2008.It is alleged that the plaintiff was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4).The event is currently under investigation.A supplemental report will be provided upon conclusion.
 
Event Description
This additional information received on 02/jan/2018 as follows: patient received an implant on (b)(6) 2008 due to unable to be anticoagulated.Patient is alleging shortness of breath, post implant pe, pain, gi and back problems due to the device.Device was successfully retrieved on (b)(6) 2013 via open, surgical removal.
 
Manufacturer Narrative
Investigation ¿ the following allegations have been investigated: suffering, mental anguish, loss of enjoyment of life¿ and limited physical activity, shortness of breath.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported suffering, mental anguish, loss of enjoyment of life¿ and limited physical activity, shortness of breath are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Reason for implant is ¿motor vehicle accident¿.Patient is further alleging ¿posterior ivc filter leg fractured into the disk space posteriorly, bilateral thrombectomy of iliac venous system and inferior vena cava, breakthrough dvt and thrombus within the iliac systems in the vena cava.¿ patient further alleges ¿suffering, mental anguish, loss of enjoyment of life¿ and limited physical activity.
 
Manufacturer Narrative
Investigation ¿ investigation is reopened due to additional information provided.The following allegations have been investigated: thrombus-deep vein thrombosis(dvt).The reported allegations have been further investigated based on the information provided to date.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
No additional information provided at this time.
 
Manufacturer Narrative
It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating, "vc and organ perforation, embedment, fracture, pe, sob, open surgery to remove, pain, inflammatory response." cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae, e.G., filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter fracture is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Fracture of a filter leg can be due to repetitive motion on a filter leg in an unusual stressed position.Among other causes, filter fracture may be associated with a filter leg perforating the ivc, a filter leg being caught in a side branch (e.G.Renal vein), excessive force or manipulations near an implanted filter (e.G.A surgical procedure in the vicinity of a filter) and / or procedures that involve other devices being passed through an in situ filter.It has been reported that retrieval of a fractured filter or filter fragments using endovascular techniques is possible.Fracture of the wire is a known risk in relation to an implanted filter and reported in the published scientific literature.It is known from the published scientific literature that a filter fragment embolized into the heart or lung may be safely retrieved.A filter that is embedded in the wall of the ivc may be difficult to retrieve.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Pe is a known risk in relation to filter implant reported in the published scientific literature.Also, it is reported that the pulmonary embolism in some cases may originate from upper extremities instead of lower extremity veins.With all filters, there is some risk of further pulmonary embolism.Unknown if the reported pain and inflammatory response is directly related to the filter and unable to identify corresponding failure mode(s) at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
This additional information received on 02jan2018 as follows: patient [pt] received an implant on (b)(6) 2008 due to unable to be anticoagulated, severe brain injury.[pt] is alleging vena cava perforation, protrusion of multiple filter legs, with significant inflammatory response around duodenum and lumbar vertebral bodies, exploration of ivc, ivc perforation repair.[pt] further alleges pain, back problems, gastrointestinal problems, clotting issues, the need for additional surgery, shortness of breath and affects the ability to be active.Device was successfully retrieved on (b)(6) 2013 via open, surgical removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUNTHER TULIP FEMORAL VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7125136
MDR Text Key95177195
Report Number1820334-2017-04444
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K043509
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/01/2011
Device Catalogue NumberIGTCFS-65-FEM
Device Lot Number2082981
Was Device Available for Evaluation? No
Date Manufacturer Received08/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other; Required Intervention;
Patient Weight104
-
-