Catalog Number 9735669
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: a medtronic representative went to the site to test the equipment. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional. The issue could not be duplicated. No parts have returned to the manufacturer for evaluation.
 
Manufacturer Narrative
No parts have been received by the manufacturer for evaluation. Part not returned for evaluation.
 
Event Description
A medtronic representative reported that during functional endoscopic sinus surgery (fess), an inaccuracy of 4 millimeters anterior was observed after navigating accurately for some time. It was reported that multiple instruments were accurate initially however the same instruments became inaccurate by the same amount. The accuracy was confirmed by the surgeon by confirming between the two front teeth at the beginning of the procedure. When the inaccuracy was suspected, surgeon went back to same spot and was inaccurate. Surgeon proceeded with surgery. The procedure was completed with the use of navigation. There was no delay to procedure. No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S8 EM ENT SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stephanie riley
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7126161
MDR Text Key103040038
Report Number1723170-2017-05189
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 01/12/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number9735669
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-