Catalog Number 1014256-040
Device Problems Difficult to Insert (1316); Difficult to Remove (1528); Folded (2630); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Evaluation summary: visual and functional inspections were performed on the returned device. The reported difficult to insert, wrinkled balloon and shaft was confirmed. The reported difficult to remove was unable to be confirmed due to device condition. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no other similar incidents reported from this lot. The investigation determined the reported difficulties appear to be related to circumstances of the procedure. Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Manufacturer Narrative
(b)(4). Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information. The device is expected to be returned for evaluation. It has not yet been received. A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that the procedure was to treat a moderately calcified lesion in the superficial femoral artery. A 5. 0x40mm armada 18 balloon dilatation catheter was used for pre-dilatation; however, minimal force was needed when it was removed from the 4f-sheath. An attempt to re-insert the balloon catheter was made; however, it was impossible due to the proximal end of the balloon and the shaft of the balloon catheter proximal to the balloon were wrinkled. There was no clinically significant delay in the procedure and no adverse patient effects. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 18 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7126558
MDR Text Key128309038
Report Number2024168-2017-09830
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K151317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 01/30/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/31/2020
Device Catalogue Number1014256-040
Device Lot Number7072041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-