LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problems Failure to Pump (1502); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 11/23/2017
Event Type  malfunction  
Manufacturer Narrative
The replaced pump was returned to livanova (b)(4) for further investigation. During evaluation, the reported issue was confirmed. It was found that the pump was very dirty and the rods were found to be rusted. Bearing damage was also identified in the motor. These issues were determined to be the cause for the reported issue.
 
Event Description
See initial report.
 
Manufacturer Narrative
There was no patient involvement. The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and found the pump difficult to manually turn. The technician was able to unblock the pump and start the device. However, it now produced noise that sounded like a broken bearing. The pump was replaced and subsequent functional verification testing was completed without further issues. The unit was returned to service. The replaced pump was requested for return to livanova (b)(4) for further evaluation. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error message on the patient side during priming. The pump did not start. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7126962
MDR Text Key258434223
Report Number9611109-2017-01028
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeBE
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/19/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-