Model Number 9735665
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2017
Event Type  malfunction  
Manufacturer Narrative
The logs for the navigation system were reviewed by medtronic personnel. However, the logs provided no additional insight into the probable cause of the anomaly. The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided.
 
Manufacturer Narrative
Software analysis reported that this occurrence will not be added to the existing test track record (b)(4) as that record was closed with the release of stealthstation s8 application version 1. 2. 0 which incorporated the fix for this anomaly. If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A medtronic representative went to site to test the equipment. Representative was unable to replicate the reported issue. A system checkout was performed and the hardware, software, and instruments passed the system checkout. The system was found to be fully functional. No parts were replaced. No parts have been received by the manufacturer for evaluation.
 
Event Description
A medtronic representative reported that during cranial resection, a low performance error was displayed in the navigation system. It was reported that exiting and resuming patient resolved the issue. The procedure was completed with the use of navigation. There was a delay of 30 seconds. No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S8 PREMIUM SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7130970
MDR Text Key106444075
Report Number1723170-2017-05182
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 08/28/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-