• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1120350-15
Device Problems Difficult to Remove (1528); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a de novo lesion in the mildly tortuous, 70% stenosed, left anterior descending (lad) coronary artery.A 3.50 x 15 mm xience alpine stent delivery system (sds) was advanced with no resistance felt to the lesion and it was observed under fluoroscopy that the stent implant was crumpled (damaged) on one side.The sds was removed from the anatomy with a bit of resistance felt and the damage to the stent implant was confirmed.A new unspecified sds was used to successfully complete the procedure.There was no reported clinically significant delay in the procedure.There was no reported adverse patient sequela.No additional information provided.
 
Manufacturer Narrative
(b)(4).The device was returned for analysis.The reported crushed stent (material deformation) was confirmed.The reported difficulty to remove was unable to be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the stent was damaged due to an interaction with the anatomy during positioning in the lesion causing the reported material deformation.Once the stent damage was identified the device was removed and likely interacted with guiding catheter during retraction causing the reported difficulty to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7132136
MDR Text Key95522545
Report Number2024168-2017-09952
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/10/2020
Device Catalogue Number1120350-15
Device Lot Number7080241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-