Catalog Number UNKNOWN |
Device Problems
Unintended Movement (3026); Insufficient Information (3190)
|
Patient Problems
Internal Organ Perforation (1987); Pain (1994); Perforation of Vessels (2135); Anxiety (2328); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Catalog# is unknown but referred to as cook gunther tulip filter.Name and address for importer site: (b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook gunther tulip filter on (b)(6) 2012".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
|
Event Description
|
This additional information received on 09feb2018 as follows: pt allegedly received an implant on (b)(6) 2012 via the right internal jugular vein due to a combination of deep vein thrombosis and pulmonary embolism.Pt.Is alleging tilt, mesenteric perforation and embedment.Additionally pt.Alleges pain and anxiety.The filter is reported to have been removed on (b)(6) 2018.
|
|
Manufacturer Narrative
|
(b)(4).Name and address for importer site: (b)(4).Additional information: investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tilt, mesenteric perforation, embedment'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.A filter that is embedded in the wall of the ivc may be difficult to retrieve.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
(b)(4).Blank fields on this form indicate the information is unknown or unavailable, or unchanged.(b)(6).Investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿tulip - tilt, mesenteric (organ) perforation, embedment, diff to ret, pain and anxiety".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.A filter that is embedded in the wall of the ivc may be difficult to retrieve.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Unknown if the reported pain is directly related to the filter.Unknown if the reported anxiety is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Filter tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
(b)(4) this report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
|
Event Description
|
This additional information received on 24may2018 as follows: pt.Further alleges vena cava perforation, mental anguish and stress.
|
|
Event Description
|
No additional information provided at this time.
|
|
Manufacturer Narrative
|
Additional information: investigation ¿ investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Unknown if the reported stress, anguish are directly related to the filter and unable to identify a corresponding failure mode at this point in time.The following allegations have been investigated: stress, anguish.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.The catalog number and lot number are unknown; no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).G1) name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.
|
|
Event Description
|
Successful filter retrieval performed on (b)(6) 2018.
|
|
Manufacturer Narrative
|
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information provided at this time.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(4).Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿tilt, vc & organ perforation, embedment, difficult to retrieve, pain and anxiety".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.A filter that is embedded in the wall of the ivc may be difficult to retrieve.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Unknown if the reported pain is directly related to the filter.Unknown if the reported anxiety is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot# are unknown, but filter tulip is manufactured and inspected according to manufacturing instructions.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|