Catalog Number UNKNOWN |
Device Problems
Unintended Movement (3026); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Internal Organ Perforation (1987); Pain (1994); Thrombosis (2100); Perforation of Vessels (2135); No Information (3190)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Catalog# is unknown but referred to as cook gunther tulip filter.Name and address for importer site: (b)(4).Since catalog# is unknown 510(k) could be either k090140, k112119 or k121057.(b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
|
|
Event Description
|
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook gunther tulip filter on (b)(6) 2013".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).Additional information: investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿tilt, vc & organ perforation, dvt, pain, nausea, anxiety, mental anguish/stress".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported pain is directly related to the filter.Unknown if the reported dvt, nausea, anxiety, mental anguish, stress are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Event Description
|
No additional information or changes at this time.
|
|
Event Description
|
This additional information received on 26feb2018 as follows: pt allegedly received an implant on (b)(6) 2013 via the right common femoral vein as prophylaxis for deep vein thrombosis (dvt) and pulmonary embolism (pe) prior to total hip arthroplasty.Pt is alleging tilt, and organ perforation (aortic, mesentery and small bowel).Pt if further alleging post implant dvt, lower abdomen and back pain, nausea and anxiety.
|
|
Manufacturer Narrative
|
Exemption number e2016032.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
|
|
Manufacturer Narrative
|
Exemption number (b)(4) lot.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Mfr site: name and address for importer site: cook medical incorporated (cmi) 400 daniels way bloomington, in 47404 registration no.: (b)(4).Additional information: investigation.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿tilt, organ perforation (aortic, mesentery, small bowel), dvt, pain, nausea, anxiety".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported peripheral vein thrombosis is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Unknown if the reported pain, nausea, and anxiety are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.H3 other text : blank fields on this form indicate the information is unknown or unavailable, or unchanged.Exemption number (b)(4) lot.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Mfr site: name and address for importer site: cook medical incorporated (cmi) 400 daniels way bloomington, in 47404 registration no.: (b)(4).Additional information: investigation.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿tilt, organ perforation (aortic, mesentery, small bowel), dvt, pain, nausea, anxiety".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported peripheral vein thrombosis is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Unknown if the reported pain, nausea, and anxiety are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
This additional information received on (b)(6) 2018 as follows: pt is additionally alleging vena cava perforation.Pt further alleges, mental anguish/stress.
|
|
Manufacturer Narrative
|
Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Patient code: vessels, perforation of (2135), listed in ifu.Name and address for importer site: (b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Patient code: vessels, perforation of (2135), listed in ifu.Name and address for importer site: (b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|