While evaluating electrode belt sn (b)(4) for an unrelated issue, a reportable problem was found.The cable connecting the distribution node and rear therapy electrodes was pulled from the strain relief.The belt was last used by a patient who passed away while using the lifevest.The patient was in asystole, which is considered a non-shockable rhythm.There is no indication that the damaged belt contributed to the patient passing.The damaged is consistent with damage observed during the removal of the device during resuscitation efforts.
|