Catalog Number 0998-00-0800-53
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/27/2017
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) is not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date. A getinge field service engineer (fse) was dispatched and performed the pneumatic interface module (pim), all manifolds, fill manifold, and drive manifold leak tests. The iabp passed all tests and the fse verified the iabp calibration, and performed the functional and safety tests to factory specifications. The fse was unable to reproduce any failures and returned the iabp to the customer cleared for clinical use.
 
Event Description
The customer reported that while the cardiosave intra-aortic balloon pump (iabp) was in use on a patient it alarmed with an autofill failure alarm. No adverse event has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7135349
MDR Text Key95595240
Report Number2249723-2017-01040
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial
Report Date 12/20/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No