• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number I7
Device Problems Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/22/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.The representative reported that the navigation system would intermittently display that the localizer was not connected.Replacement parts were shipped to the site, however confirmation of replacement has not been provided.
 
Event Description
A medtronic representative reported that, while outside of a procedure, the navigation system displayed that the localizer was not connected.The reported issue occurred between cases.It was noted that restarting the navigation system restored functionality for later procedures.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
The positioning sensor unit (psu) for the navigation system was received by the manufacturer for evaluation.Testing found that the psu intermittently tracked along the far left edge and back edge of the field of view.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Manufacturer Narrative
The in/out hub was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.The software investigation found that the reported event was unrelated to a software issue.The software functioned as designed.
 
Manufacturer Narrative
The software investigation found that the reported event was unrelated to a software issue.The software functioned as designed.
 
Manufacturer Narrative
Correction: a medtronic representative went to the site to test the equipment.Following replacement of the components, the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.
 
Manufacturer Narrative
Correction: the reported issue with the positioning sensor unit (psu) was not recorded in the medtronic navigation hardware anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7137849
MDR Text Key95503115
Report Number1723170-2017-05228
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994653482
UDI-Public00613994653482
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberI7
Device Catalogue Number9734060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/23/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-