• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0998-00-3023-53
Device Problem Inability to Auto-Fill (1044)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2017
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per getinge standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and was unable to duplicate the reported problem.The fse reviewed the error logs and noted that the device had failed autofill on several dates.The fse performed an autofill calibration and the device passed at 105 mmhg.A safety disk leak test was performed and device passed.The fse found iab fill port luer to be a suspected problem due to plastic threads being completely engaged with metal fitting on safety disk.The luer fitting was replaced and multiple autofill cycles were performed with no errors.The device ran on simulator for one hour with no errors.The iabp was returned to clinical usage.
 
Event Description
It was reported that an autofill failure occurred on the cs300 intra-aortic balloon pump (iabp) during use on a patient.There was no patient harm.No adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7138150
MDR Text Key95929371
Report Number2249723-2017-01042
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-