Catalog Number NTSED-024115-UDH
Device Problem Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event Description
The user facility reported during a ureteroscopy procedure, after multiple use throughout the procedure, the ncircle delta wire tipless stone extractor would no longer retract entirely.The procedure was able to be completed with the device.No unintended portion of the device remained inside the patient¿s body.No additional procedures were required due to this occurrence.No adverse effects or consequences were reported.
 
Manufacturer Narrative
A review of the manufacturing instructions, quality control, specifications, drawing, functional test and visual inspection of the actual device was completed during the investigation.One device was returned for investigation.The collet knob is tight and secure.The mlla (male luer lock adaptor) is loose.The pett measures 3 cm in length.A functional test determined the handle does not actuate the basket formation.The support sheath and basket sheath are still adhered.A kink was noted in the basket sheath at 85 cm from the distal tip.A visual examination noted the support sheath is bowed in appearance.After manually checking the basket sheath for kinks, damage, and anomalies, the handle actuated the basket formation.The handle was reset and reassembled and the basket actuated fine.No patient harm was reported.There is no indication that a design or process related failure mode contributed to this event.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.Based on the investigation evaluation, there is no indication that a design or process related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality; device integrity prior to shipping.Based on the provided information a definitive root cause cannot be established or reported at this time.Appropriate personnel have been notified to monitor for similar complaints.Per a quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NCIRCLE DELTA WIRE TIPLESS STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key7138364
MDR Text Key95611560
Report Number1820334-2017-04334
Device Sequence Number1
Product Code FFL
UDI-Device Identifier00827002191105
UDI-Public(01)00827002191105(17)200830(10)8182060
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberNTSED-024115-UDH
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1