LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5 (CP5); CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 60-01-04 |
Device Problem
No Flow (2991)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump 5 (cp5) the incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.Visual inspection did not identify any abnormalities or defects and several hours of troubleshooting was unable to duplicate the reported issue.The user was not comfortable returning the pump to service, so it was replaced and is planned to be returned to livanova for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Evaluated on site by livanova technician.
|
|
Event Description
|
Livanova (b)(4) received a report that the centrifugal pump 5 (cp5) displayed a pressure error and forward flow stopped during a case, resulting in a negative flow alarm.While troubleshooting the system circuit and other parts of the oxygenator, the centrifugal pump began forward flow again.There was no report of patient injury.
|
|