MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX 160NRE DIALYZER FINISHED ASSY. DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX 160NRE DIALYZER FINISHED ASSY. DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 0500316E
Device Problem Fluid/Blood Leak (1250)
Patient Problem Blood Loss (2597)
Event Date 12/07/2017
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: although it was originally stated that the dialyzer was available to be returned, to date no sample have been returned to the manufacturer for physical evaluation. The reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation. A production records review was performed on the reported lot. An investigation of the device history records (dhr) was conducted by the manufacturer. There was no indication of product non-acceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event. The lot met all release criteria. A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device. Therefore, the complaint is not confirmed. Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
A supplemental mdr will be submitted upon device evaluation.
 
Event Description
Hemodialysis nurse from a user facility reported a blood leak alarm during treatment. Blood was noted in the effluent and dialysate compartment. Blood was not returned to the patient. During follow up the nurse reported the patient was beginning treatment when the machine alarmed. Treatment was stopped. The blood was not returned, the patient lost approximately 250 ml. The patient was moved to another machine to finish treatment. There was no medical intervention for the blood loss. Test strips were used and tested positive. The dialyzer may be made available for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameOPTIFLUX 160NRE DIALYZER FINISHED ASSY.
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
475 west 13th street
ogden UT 84404
Manufacturer (Section G)
OGDEN MANUFACTURING PLANT
475 west 13th street
ogden UT 84404
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key7139774
MDR Text Key252231162
Report Number1713747-2017-00405
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100149
UDI-Public00840861100149
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K002761
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/31/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/30/2020
Device Catalogue Number0500316E
Device Lot Number17HU06011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device AgeMO
Event Location No Information
Date Manufacturer Received05/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/10/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-