MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-02H
Device Problem Insufficient Information (3190)
Patient Problems Headache (1880); Dizziness (2194)
Event Date 11/02/2017
Event Type  Injury  
Event Description
The recipient is reportedly experiencing dizziness and headaches at the implant site.The recipient's device was explanted.The recipient was not reimplanted.
 
Manufacturer Narrative
(b)(4).The recipient is reportedly doing well.The external visual inspection revealed cut silastic overmold near the antenna coil and at the fantail region.In addition, slight dents were observed on the bottom cover.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
delilah garcia
28515 westinghouse place
valencia, CA 91355
6613624711
MDR Report Key7140971
MDR Text Key95586007
Report Number3006556115-2017-00690
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/30/2008
Device Model NumberCI-1400-02H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2017
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age15 YR
-
-