• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Complete Blockage (1094); Difficult to Remove (1528); Insufficient Information (3190)
Patient Problem No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Exemption number e2016032e lot. (b)(4). This report includes information known at this time. A follow up report will be submitted should additional relevant information become available.
 
Event Description
This additional information received on 06feb2018 as follows: pt allegedly received an implant on (b)(6) 2008 via the right common femoral vein due to history of deep vein thrombosis/pulmonary embolism. Pt is alleging device is unable to be retrieved. Additionally pt. Is alleging that due to a thrombosed ivc filter, stents were placed within the inferior vena cava. Pt. States was informed by md that following this procedure that the ivc filter could not be removed.
 
Manufacturer Narrative
Exemption number e2016032. (b)(4). Investigation - it has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'celect, thrombosed ivc filter (occlusion), unable to retrieve'. Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c. F. R. 803. 56. Filter retrieval is occasionally difficult. This is well-known from published scientific literature where filter retrievals are referred to as simple vs. Complex. Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques. Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature. Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic. No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e. G. Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Manufacturer Narrative
(b)(4). Catalog# is unknown but referred to as cook celect filter. Name and address for importer site: (b)(4). Since catalog# is unknown 510(k) could be either k061815 or k073374. (b)(4). It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date. Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c. F. R. 803. 56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2008". Patient outcome: it is alleged that [pt] was injured without further explanation. Hospital and medical records have been requested but not yet provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN
Type of DeviceDTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA 4632
Manufacturer Contact
thomas hessner kirk
sandet 6
bjaeverskov 4632 4632
DA   4632
56868686
MDR Report Key7141449
MDR Text Key192552125
Report Number3002808486-2017-02508
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup
Report Date 03/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Distributor Facility Aware Date12/06/2017
Event Location No Information
Date Manufacturer Received03/16/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/22/2017 Patient Sequence Number: 1
-
-