Model Number SONATA |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
Failure of Implant (1924)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The patient has a decreasing performance with the device.Re-implantation considered.
|
|
Manufacturer Narrative
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However to determine an exact root cause a device investigation of the explanted device is necessary.A date for re-implantation has not been made available yet.
|
|
Event Description
|
The recipient has a decreasing performance with the device.No trauma has been reported.Re-implantation is considered but no date has been scheduled yet.
|
|