MAQUET CARDIOPULMONARY AG MAQUET HCU40; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HEATER-COOLER UNIT
Device Problems Fire (1245); Smoking (1585)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
It was reported that the hcu40 smells burnt.So the device was exchanged.During troubleshooting by the technician the heater of the patient circuit starts to smoke.When trying to drain the water out of the machine, the thermal switch caught fire.The technician turns the hcu off immediately and extinguishes the fire.There were no negative consequences to the patient.(b)(4).
 
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # (b)(4).(b)(6).The device was sent to rastatt for investigation and repair.According to the lce investigation report: there could have been a leak in the old main heater.(unfortunately, this is no longer reproducible, since the old main heater can no longer be installed and would no longer function), or the corrosion damage is caused by condensation that has formed around the thermal switch.The damage was due to a leak or condensation on the main heater.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.A supplemental medwatch will be submitted if new information has been received.
 
Event Description
Internal reference: (b)(4).Customer reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU40
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7142861
MDR Text Key95761949
Report Number8010762-2017-00408
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHEATER-COOLER UNIT
Device Catalogue Number701044054
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2017
Date Device Manufactured06/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-