Catalog Number 0998-00-3023-53
Device Problems Material Rupture (1546); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/24/2017
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) is not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date. A getinge field service engineer (fse) was dispatched and the customer stated the intra-aortic balloon (iab) ruptured. The fse tested the iabp and observed no evidence of blood in the system. The fse viewed the fault logs and observed no fault codes that pertain to blood being detected. The fse could not duplicate any failures and completed the checkout and returned the iabp back to the customer cleared for clinical use.
 
Event Description
The customer originally reported that the cs300 intra-aortic balloon pump (iabp) was not functional. A getinge field service engineer (fse) was dispatched and the customer told them that there was a balloon rupture while the iabp was in use on a patient. No adverse event has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7144043
MDR Text Key95923341
Report Number2249723-2017-01061
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial
Report Date 12/22/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-