Model Number CI-1400-01 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Headache (1880)
|
Event Date 05/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Udi number: na.
|
|
Event Description
|
The recipient is reportedly experiencing headaches.Revision surgery is under consideration.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
The external visual inspection revealed the electrode was cut prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed cut electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.This is an interim report.
|
|
Manufacturer Narrative
|
The external visual inspection revealed the electrode was cut prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection confirmed cut electrode wires.This is believed to have occurred during revision surgery.System lock was verified.The electrode condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.This device had moisture that exceeded the residual gas analysis test limit.Leak testing did not reveal any leaks.Corrective actions were implemented.This is the final report.
|
|
Search Alerts/Recalls
|