Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was missing from clevis approximately 0.032 x 0.046 in size in the clevis.The clevis did not exhibit any damage or wear marks.Other cables at the wrist were not damaged.Isi has conducted a device history record (dhr) review for this device and did not find any non-conformances that were related to this reported event.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis suggests that if the malfunction were to recur it could cause or contribute to an adverse event.
|
It was reported that during a da vinci-assisted thoracic sympathectomy procedure, a snapped cable was noted on the small grasping retractor instrument.There was no report of fragment(s) falling inside the patient, patient harm, adverse outcome or injury.
|