ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problem Folded (2630)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2017
Event Type  malfunction  
Manufacturer Narrative
Evaluation summary: the device was not returned for evaluation.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.Additional information was requested.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported while using a preloaded intraocular lens (iol) delivery system, the lens folded out in the incision and the lens was removed.A backup lens was used to complete the procedure.Additional information was requested.
 
Manufacturer Narrative
The device was returned in loose in a plastic bag.The plunger lock and lens stop have been removed.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger has been retracted just behind the wound guard.No damage or abnormalities are observed.The lens was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The root cause for the reported "iol folded out in the incision" could not be determined.No problem was found with the returned device.It is important to keep the wound guard securely against the eye and fully advance the plunger for proper lens delivery.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7145254
MDR Text Key95820810
Report Number1119421-2017-01468
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model NumberAU00T0
Device Lot Number12501621
Other Device ID Number00380652358408
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2018
Date Device Manufactured10/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-