Model Number S7
Device Problems Poor Quality Image (1408); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that a video graphics array (vga) splitter was replaced which resolved the issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect video graphics array (vga) has not been received by the manufacturer for evaluation.
 
Event Description
A site representative reported that while outside of procedure, the image on both the monitors of navigation system was bad and the video signal was scrolling horizontally.There was no patient present at the time of the issue.
 
Manufacturer Narrative
The suspect splitter has been received by the manufacture for evaluation.When the returned part was connected to a known good system the splitter returned good image for each output.No problem found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7145735
MDR Text Key95832145
Report Number1723170-2017-05289
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169505742
UDI-Public00643169505742
Combination Product (y/n)N
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-