MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
 
Event Description
A site representative reported that during spinal fusion procedure, the site was unable to push images from imaging system to navigation system.Multiple attempts were made to push exams and the imaging system and navigation system were rebooted several times but the issue was not resolved.The procedure was completed without the use of navigation.There was a delay of less than 1 hour.No known impact on patient outcome.
 
Manufacturer Narrative
Three documented attempts were made to obtain the software logs from the site.All three attempts were unsuccessful.No further action required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7146016
MDR Text Key95848022
Report Number1723170-2017-05293
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age61 YR
Patient Weight87
-
-