MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CART 9734056 S7 STAFF SHRT 100-120V INTL SEE H10)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) CART 9734056 S7 STAFF SHRT 100-120V INTL SEE H10) Back to Search Results
Model Number S7
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
No procode, common device name and/or 510k provided as this device is not released for distribution in the united states.
 
Manufacturer Narrative
The computer for the navigation system returned to the manufacturer for analysis. The device was found to be fully functional with no problem found. The reported event could not be duplicated by medtronic personnel.
 
Manufacturer Narrative
Patient information not provided due to japanese patient privacy regulations. The software investigation found that the reported event was related to a software issue. This issue was documented in a medtronic navigation software anomaly tracking database.
 
Event Description
A site representative reported that during a procedure, the navigation system unexpectedly rebooted without any prompt form user and upon booting up a "no signal" was displayed. Rebooting the system multiple times resolved the issue and the site proceeded with procedure. The procedure was completed with the use of navigation. There was no delay to procedure. No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCART 9734056 S7 STAFF SHRT 100-120V INTL
Type of DeviceSEE H10)
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7146324
MDR Text Key103050777
Report Number1723170-2017-05297
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
SEE H10)
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 02/13/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9734056
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/06/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-