Model Number S7 |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/30/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
No procode, common device name and/or 510k provided as this device is not released for distribution in the united states.
|
|
Manufacturer Narrative
|
The computer for the navigation system returned to the manufacturer for analysis.
The device was found to be fully functional with no problem found.
The reported event could not be duplicated by medtronic personnel.
|
|
Manufacturer Narrative
|
Patient information not provided due to japanese patient privacy regulations.
The software investigation found that the reported event was related to a software issue.
This issue was documented in a medtronic navigation software anomaly tracking database.
|
|
Event Description
|
A site representative reported that during a procedure, the navigation system unexpectedly rebooted without any prompt form user and upon booting up a "no signal" was displayed.
Rebooting the system multiple times resolved the issue and the site proceeded with procedure.
The procedure was completed with the use of navigation.
There was no delay to procedure.
No impact on patient outcome.
|
|