• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint rt380 circuit was returned to fisher & paykel healthcare in (b)(4) for investigation.The returned device was visually inspected.Results: visual inspection revealed that the collar at the patient end of the expiratory limb was cracked along its length.Conclusion: we were unable to determine conclusively what had caused the collar to crack; however, based on our previous investigations it is possible that the crack is due to being in contact with a chemical solution, resulting in environmental stress cracking.As the damage occurred after 13 days of use, we can conclude that the circuit was originally functioning correctly.All rt380 adult dual-heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The user instructions that accompany the rt380 breathing circuit state: - do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.-perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.-set appropriate ventilator alarms.
 
Event Description
A healthcare facility in (b)(6) reported via a distributor that they found a crack on the collar of the expiratory limb of an rt380 evaqua2 adult breathing circuit after 13 days of patient use.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
sharon kuruvilla
173 technology drive
suite 100
irvine, CA 92618-2216
9494534000
MDR Report Key7147022
MDR Text Key96021597
Report Number9611451-2017-01245
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Was the Report Sent to FDA? No
Date Manufacturer Received11/30/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-