• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AVANCE; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AVANCE; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2017
Event Type  malfunction  
Manufacturer Narrative
The distributor performed a checkout of the equipment and confirmed the reported complaint.The anesthesia control board was replaced.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported that, during boot-up, the unit went into a system malfunction.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7147546
MDR Text Key95855753
Report Number2112667-2017-02397
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K032803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial
Report Date 12/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/01/2017
Date Device Manufactured06/04/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-