Model Number S7 |
Device Problems
Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information not provided due to (b)(6) patient privacy regulations.A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
A site representative reported that during a procedure, the camera of the navigation system could not verify active frame and scope probe.It was reported that the navigation system and software were functioning normally.The active frame was not recognized indicating "not connected".System was rebooted which resolved the issue however, the issue occurred again after a while.Rebooting the system again resolved the issue and the site proceeded with case.The procedure was completed with the use of navigation.There was no delay to procedure.No impact on patient outcome.
|
|
Manufacturer Narrative
|
The polaris spectra system control unit (scu) was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
|
Manufacturer Narrative
|
No procode, common device name, unique device identification (udi) and/or 510k provided as this device is not released for distribution in the united states.Correction: product, unique device identification (udi), device manufacture date and related fields updated to proper value.
|
|