• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND, INC S8-3T MICRO TEE TRANSDUCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND, INC S8-3T MICRO TEE TRANSDUCER Back to Search Results
Lot Number B0L3FK
Device Problem Poor Quality Image (1408)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event Description
A customer reported an s8-3t model transducer was producing poor image quality.There was no injury associated with this event.
 
Manufacturer Narrative
Evaluation of this transducer confirmed poor image quality as described by the customer.Investigation of the device noted exposure to fluid in the connector, protein buildup on the bead and tip shell, and damage to the connector and control handle.Further investigation revealed stains and/or discoloration of labels, brake and cable jacket as well as damage to the locking latch and steering mechanism.The extensive damage identified during the evaluation would inhibit the overall performance of this transducer and is indicative of improper maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S8-3T MICRO TEE TRANSDUCER
Type of Device
S8-3T MICRO TEE TRANSDUCER
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
paul corrigan
22100 bothell everett highway
bothell, WA 98021
4254877000
MDR Report Key7149452
MDR Text Key95951352
Report Number3019216-2017-00045
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K030455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot NumberB0L3FK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/21/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-