• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Internal Organ Perforation (1987); Perforation of Vessels (2135); Anxiety (2328); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Catalog# is unknown but referred to as cook gunther tulip filter.Name and address for importer site: (b)(4).Since catalog# is unknown 510(k) could be either k090140, k112119 or k121057.(b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook gunther tulip filter on (b)(6) 2014".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
Exemption number e2016032.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event Description
This additional information received on 01feb2018 as follows: pt allegedly received an implant on (b)(6) 2014 via the right femoral vein as prophylactic.Pt is alleging organ perforation (small bowel), vena cava perforation.Pt further alleges anxiety.
 
Manufacturer Narrative
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(4).Additional information: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'organ perforation (small bowel), vena cava perforation, anxiety'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported anxiety is directly related to the filter and unable to identify corresponding failure mode(s) at this time.Rpn and lot# are unknown, but the tulip filter is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Blank fields on this form indicate the information is unknown or unavailable, or unchanged.Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(6).Additional information: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'organ perforation (small bowel), vena cava perforation, anxiety'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported anxiety is directly related to the filter and unable to identify corresponding failure mode(s) at this time.Rpn and lot# are unknown, but the tulip filter is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Blanks on this form indicate the information is unknown or unavailable, or unchanged.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.H3 other text : blank fields on this form indicate the information is unknown or unavailable, or unchanged.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
This additional information received on 11apr2018 as follows: there have been no reported attempts at filter retrieval.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key7150508
MDR Text Key95897398
Report Number3002808486-2017-02523
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup,Followup
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Distributor Facility Aware Date12/13/2017
Date Manufacturer Received04/11/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight64
-
-