• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. MODULAR HANDLE, T NON-RATCHETING TORQUE LIMITING 90 IN-LB; SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. MODULAR HANDLE, T NON-RATCHETING TORQUE LIMITING 90 IN-LB; SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM Back to Search Results
Catalog Number 3356-2
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/02/2017
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that the torque output of a torque limiting handle was found outside of the adequate performance range.This was detected during a preliminary calibration check by zimmer biomet spine personnel.There are no specific surgical procedures associated with this handle.
 
Manufacturer Narrative
The handle was not returned for evaluation, so no results are available and no conclusions can be drawn.A review of the dhr did not reveal any manufacturing related issues which would have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR HANDLE, T NON-RATCHETING TORQUE LIMITING 90 IN-LB
Type of Device
SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7150723
MDR Text Key95935028
Report Number3012447612-2017-00708
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK131980
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number3356-2
Device Lot Number86KD-012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-