MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Device Problems False Reading From Device Non-Compliance (1228); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. No conclusion can be drawn at this time. We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm. The customer¿s blood glucose was 192 mg/dl and the sensor glucose was 72 mg/dl at the time of the incident. Troubleshooting was done and customer stated that their blood glucose and sensor glucose levels were not in acceptable range. The current blood glucose was 191 mg/dl. Insulin delivery was suspended due to sensor glucose values. Sensor glucose value that triggered the suspend event was 68 mg/dl. Suspend on low limit in sensor settings was 70 mg/dl. The sensor will not be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSENSOR ENLITE MMT-7008A
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7151044
MDR Text Key96011979
Report Number2032227-2017-77694
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 12/28/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/14/2018
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG2225D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/16/2017
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-