Model Number FUSION COMPACT |
Device Problems
Device Stops Intermittently (1599); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/30/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Device lot number, or serial number, unavailable.Device manufacturing date is dependent on lot number/serial number, therefore, unavailable.No parts have been returned to the manufacturer for evaluation.
|
|
Event Description
|
A medtronic representative reported that while outside of a procedure, the axiem ® box was functioning intermittently on the navigation system and was displaying an 8 in the controller status.There was no patient present when this issue was identified.
|
|
Manufacturer Narrative
|
Product and associated fields updated to proper value.The interface box for the navigation system was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.Testing revealed that hardware did not pass testing and the axiem was replaced.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
|
|